The pyruvate-lactate axis modulates cardiac hypertrophy and heart failure Atrial Fibrillation and the Prognostic Performance of Biomarkers in Heart Failure Meta-analysis of outcomes after intravascular ultrasound-guided versus angiography-guided drug-eluting stent implantation in 26,503 patients enrolled in three randomized trials and 14 observational studies Dapagliflozin and Cardiovascular Outcomes in Type 2 Diabetes Efficacy and Safety of Dapagliflozin in Heart Failure With Reduced Ejection Fraction According to Age: Insights From DAPA-HF Age-Related Characteristics and Outcomes of Patients With Heart Failure With Preserved Ejection Fraction Mechanical circulatory support devices for acute right ventricular failure In patients with stable coronary heart disease, low-density lipoprotein-cholesterol levels < 70 mg/dL and glycosylated hemoglobin A1c < 7% are associated with lower major cardiovascular events Mechanical circulatory support devices in advanced heart failure: 2020 and beyond Pulmonary artery denervation to treat pulmonary arterial hypertension: the single-center, prospective, first-in-man PADN-1 study (first-in-man pulmonary artery denervation for treatment of pulmonary artery hypertension)

Original Research2018 Feb;27(2):212-218.

JOURNAL:Heart Lung Circ. Article Link

The Utility of Contrast Medium Fractional Flow Reserve in Functional Assessment Of Coronary Disease in Daily Practice

Van Wyk P, Puri A, Blake J et al. Keywords: Contrast Fractional Flow Reserve

ABSTRACT


BACKGROUND Adenosine induced hyperaemic fractional flow reserve (aFFR) is a validated predictor of clinical outcome and part of routine interventional practice. Protocol issues associated with the adenosine infusion limit the use of aFFR in clinical practice. Contrast medium induced hyperaemic FFR (cFFR) is a simpler procedure from a practical standpoint. We compared the two in a real world setting.


METHODS - We analysed 76 patients that had both cFFR and aFFR assessment of 100 angiographically indeterminate coronary stenosis. cFFR was performed with intracoronary contrast medium injections (10ml for left coronary lesions and 8ml for right coronary lesions). The diagnostic performance of cFFR was analysed and compared to the gold standard aFFR.


RESULTS Mean cFFR was 0.87 (±0.07) and mean aFFR was 0.84 (±0.08). Bland-Altman analysis revealed a close agreement between cFFR and aFFR (0.035±0.032; 95% CI: -0.028 to 0.098) and good linear correlation (r=0.92, r2=0.86; p<0.0001). Using cFFR cut-off values of ≤0.83 in predicting an aFFR value of ≤0.80 or a cFFR value ≥0.88, predicting an aFFR value of >0.80 yielded a sensitivity of 100%, specificity of 96.1%, positive predictive value of 92.3%, negative predictive value of 100% and diagnostic accuracy of 96%. Only 24% of cFFR values were in the 0.84 to 0.87 range.


CONCLUSION - Contrast medium induced hyperaemic FFR as an initial assessment may limit the need for adenosine to when cFFR falls in the 0.84 to 0.87 range. The use of adenosine infusion potentially could have been avoided in the majority of patients in this study.