IVUS Guidance for Coronary Revascularization: When to Start, When to Stop? Prognostic implication of lipidomics in patients with coronary total occlusion undergoing PCI Incidence and Clinical Outcomes of Stent Fractures on the Basis of 6,555 Patients and 16,482 Drug-Eluting Stents From 4 Centers Left Ventricular Rapid Pacing Via the Valve Delivery Guidewire in Transcatheter Aortic Valve Implantation Significantly less inappropriate shocks in ischemic patients compared to non-ischemic patients: The S-ICD experience of a high volume single-center A Genotype-Guided Strategy for Oral P2Y12 Inhibitors in Primary PCI Intravascular ultrasound findings of early stent thrombosis after primary percutaneous intervention in acute myocardial infarction: a Harmonizing Outcomes with Revascularization and Stents in Acute Myocardial Infarction (HORIZONS-AMI) substudy Blood Pressure Assessment in Adults in Clinical Practice and Clinic-Based Research: JACC Scientific Expert Panel 3D Printing and Heart Failure: The Present and the Future Coronary bifurcation lesions treated with simple or complex stenting: 5-year survival from patient-level pooled analysis of the Nordic Bifurcation Study and the British Bifurcation Coronary Study

Clinical Trial2020 Dec 16;S1936-8798(20)32011-2.

JOURNAL:JACC Cardiovasc Interv. Article Link

Suture- or Plug-Based Large-Bore Arteriotomy Closure: A Pilot Randomized Controlled Trial

MP van Wiechen, D Tchétché, N Dumonteil et al. Keywords: TAVR; vascular closure device; dedicated plug-based VCD vs suture-based VCD

ABSTRACT


OBJECTIVES - The authors sought to test the superiority in terms of efficacy and safety of a dedicated plug-based vascular closure device (VCD) during transcatheter aortic valve replacement (TAVR) over a suture-based VCD.


BACKGROUND - Vascular complications after TAVR are relevant and often associated with VCD failure.


METHODS - The MASH trial (MANTA vs. Suture-based vascular closure after transcatHeter aortic valve replacement) is an international, 2-center pilot randomized controlled trial comparing the MANTA VCD (Teleflex, Wayne, Pennsylvania) versus 2 ProGlides (Abbott Vascular, Abbott Park, Illinois). The primary composite endpoint consisted of access siterelated major or minor vascular complications at 30-daysfollow-up. Secondary endpoints included clinically relevant access site bleeding, time to hemostasis, and modified VCD failure (defined as failure to achieve hemostasis within 5 min or requiring additional endovascular maneuvers such as endovascular stenting, surgical techniques, or additional closure devices). Adverse events were adjudicated by an independent clinical events committee according to the VARC-2 definitions.


RESULTS - A total of 210 TAVR patients were included between October 2018 and January 2020. Median age was 81 years, 54% were male, and the median STS score was 2.7%. There was no significant difference in the primary endpoint of access siterelated vascular complications between MANTA and ProGlide (10% vs. 4%; p = 0.16). Clinically significant access site bleedings were similar with both closure techniques (9% vs. 6%; p = 0.57). Modified VCD failure occurred less frequently in MANTA versus ProGlide (20% vs. 40%; p < 0.01). Suture-based closure required more often additional closure devices, whereas MANTA numerically needed more covered stents and surgical bailouts.


CONCLUSIONS - Plug-based large-bore arteriotomy closure was not superior to suture-based closure. Plug-based closure required fewer, but a different kind of bailout maneuvers.