CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

Feasibility of Coronary Access and Aortic Valve Reintervention in Low-Risk TAVR Patients Efficacy and Safety of Dapagliflozin in Heart Failure With Reduced Ejection Fraction According to Age: Insights From DAPA-HF Clinical trial design and rationale of the Multicenter Study of MagLev Technology in Patients Undergoing Mechanical Circulatory Support Therapy With HeartMate 3 (MOMENTUM 3) investigational device exemption clinical study protocol Optical Frequency Domain Imaging Versus Intravascular Ultrasound in Percutaneous Coronary Intervention (OPINION Trial) Results From the OPINION Imaging Study The year in cardiovascular medicine 2020: heart failure and cardiomyopathies Atrial Fibrillation and the Prognostic Performance of Biomarkers in Heart Failure A three-vessel virtual histology intravascular ultrasound analysis of frequency and distribution of thin-cap fibroatheromas in patients with acute coronary syndrome or stable angina pectoris H2FPEF Score for Predicting Future Heart Failure in Stable Outpatients With Cardiovascular Risk Factors Unexpectedly Low Natriuretic Peptide Levels in Patients With Heart Failure Fractional flow reserve derived from CCTA may have a prognostic role in myocardial bridging

Clinical Trial2020 Dec 16;S1936-8798(20)32011-2.

JOURNAL:JACC Cardiovasc Interv. Article Link

Suture- or Plug-Based Large-Bore Arteriotomy Closure: A Pilot Randomized Controlled Trial

MP van Wiechen, D Tchétché, N Dumonteil et al. Keywords: TAVR; vascular closure device; dedicated plug-based VCD vs suture-based VCD

ABSTRACT


OBJECTIVES - The authors sought to test the superiority in terms of efficacy and safety of a dedicated plug-based vascular closure device (VCD) during transcatheter aortic valve replacement (TAVR) over a suture-based VCD.


BACKGROUND - Vascular complications after TAVR are relevant and often associated with VCD failure.


METHODS - The MASH trial (MANTA vs. Suture-based vascular closure after transcatHeter aortic valve replacement) is an international, 2-center pilot randomized controlled trial comparing the MANTA VCD (Teleflex, Wayne, Pennsylvania) versus 2 ProGlides (Abbott Vascular, Abbott Park, Illinois). The primary composite endpoint consisted of access siterelated major or minor vascular complications at 30-daysfollow-up. Secondary endpoints included clinically relevant access site bleeding, time to hemostasis, and modified VCD failure (defined as failure to achieve hemostasis within 5 min or requiring additional endovascular maneuvers such as endovascular stenting, surgical techniques, or additional closure devices). Adverse events were adjudicated by an independent clinical events committee according to the VARC-2 definitions.


RESULTS - A total of 210 TAVR patients were included between October 2018 and January 2020. Median age was 81 years, 54% were male, and the median STS score was 2.7%. There was no significant difference in the primary endpoint of access siterelated vascular complications between MANTA and ProGlide (10% vs. 4%; p = 0.16). Clinically significant access site bleedings were similar with both closure techniques (9% vs. 6%; p = 0.57). Modified VCD failure occurred less frequently in MANTA versus ProGlide (20% vs. 40%; p < 0.01). Suture-based closure required more often additional closure devices, whereas MANTA numerically needed more covered stents and surgical bailouts.


CONCLUSIONS - Plug-based large-bore arteriotomy closure was not superior to suture-based closure. Plug-based closure required fewer, but a different kind of bailout maneuvers.