Transcatheter Aortic Valve Replacement vs Surgical Replacement in Patients With Pure Aortic Insufficiency Current treatment of significant left main coronary artery disease: A review Incidence, predictors, and outcomes associated with acute kidney injury in patients undergoing transcatheter aortic valve replacement: from the BRAVO-3 randomized trial The conductive function of biopolymer corrects myocardial scar conduction blockage and resynchronizes contraction to prevent heart failure The Role of the Pericardium in Heart Failure: Implications for Pathophysiology and Treatment Percutaneous Left Atrial Appendage Closure for Stroke Prophylaxis in Patients With Atrial Fibrillation: 2.3-Year Follow-up of the PROTECT AF (Watchman Left Atrial Appendage System for Embolic Protection in Patients With Atrial Fibrillation) Trial Coronary plaque redistribution after stent implantation is determined by lipid composition: A NIRS-IVUS analysis Prognostic implications of baseline 6‐min walk test performance in intermediate risk patients undergoing transcatheter aortic valve replacement Three vs twelve months of dual antiplatelet therapy after zotarolimus-eluting stents: the OPTIMIZE randomized trial Management of Antithrombotic Therapy in Atrial Fibrillation Patients Undergoing PCI: JACC State-of-the-Art Review

Original ResearchApr 03, 2022

JOURNAL:JACC Clin Electrophysiol Article Link

Clinical Impact of Residual Leaks Following Left Atrial Appendage Occlusion: Insights From the NCDR LAAO Registry

M Alkhouli, CA Du , A Killu et al.

ABSTRACT

BACKGROUND - Data on the impact of residual peri-device leak after left atrial appendage occlusion (LAAO) are limited.

 

OBJECTIVES - The goal of this study was to explore the association of peri-device leak with adverse clinical events.

 

METHODS - The National Cardiovascular Data Registry LAAO Registry was queried to identify patients undergoing LAAO between January 1, 2016, and December 31, 2019. Patients were classified according to leak size on echocardiography at 45 ± 14 days (0 mm, no leak; >0-5 mm, small leak; and >5 mm, large leak).

 

RESULTS - A total of 51,333 patients were included, of whom 37,696 (73.4%) had no leak, 13,258 (25.8%) had small leaks, and 379 (0.7%) had large leaks. The proportion of patients on warfarin at 45 days was higher in the large vs small or no leak cohorts (44.9% vs 34.4% and 32.4%, respectively; P < 0.001). At 6 and 12 months, anticoagulant utilization decreased but remained more frequent in patients with large leaks. Thromboembolic and bleeding events were uncommon in all groups. However, compared with patients with no leak, those with small leaks had slightly higher odds of stroke/transient ischemic attack/systemic embolization (adjusted HR: 1.152; 95% CI: 1.025-1.294), major bleeding (HR: 1.11; 95% CI: 1.029-1.120), and any major adverse events (HR: 1.102; 95% CI: 1.048-1.160). There were no significant differences in adverse events between patients with large leaks and patients with small or no leaks.

 

CONCLUSIONS - Small (>0-5 mm) leaks after LAAO were associated with a modestly higher incidence of thromboembolic and bleeding events; large leaks (>5 mm) were not associated with adverse events, although higher proportions of these patients were maintained on anticoagulation. Newer devices with improved seal might mitigate the events associated with residual leaks.