The Current State of Left Main Percutaneous Coronary Intervention Genetics and Causality of Triglyceride-Rich Lipoproteins in Atherosclerotic Cardiovascular Disease sST2 Predicts Outcome in Chronic Heart Failure Beyond NT−proBNP and High-Sensitivity Troponin T Glucose-lowering Drugs or Strategies, Atherosclerotic Cardiovascular Events, and Heart Failure in People With or at Risk of Type 2 Diabetes: An Updated Systematic Review and Meta-Analysis of Randomised Cardiovascular Outcome Trials Potential protective mechanisms of green tea polyphenol EGCG against COVID-19 Prevention of Bleeding in Patients with Atrial Fibrillation Undergoing PCI Radial Versus Femoral Access for Rotational Atherectomy: A UK Observational Study of 8622 Patients Effect of empagliflozin on exercise ability and symptoms in heart failure patients with reduced and preserved ejection fraction, with and without type 2 diabetes The outcomes of intravascular ultrasound-guided drug-eluting stent implantation among patients with complex coronary lesions: a comprehensive meta-analysis of 15 clinical trials and 8,084 patients Limitations of Repeat Revascularization as an Outcome Measure

Clinical Trial2025 Nov 24;18(22):2701-2710.

JOURNAL:JACC Cardiovasc Interv . Article Link

Paclitaxel-Coated Balloon for the Treatment of Small Vessel In-Stent Restenosis: A Subgroup Analysis of the AGENT IDE Randomized Trial

J Wen, S Dohad, R Shlofmitz et al. Keywords: drug-coated balloon; in-stent restenosis; small vessel; target lesion failure; target lesion revascularization; uncoated balloon.

Abstract

BACKGROUD -  Treatment of small vessel (SV) coronary artery disease is associated with higher restenosis rates. Drug-coated balloons offer a promising treatment option for stent failure by delivering an antiproliferative drug and avoiding an additional metal implant. However, evidence supporting the use of paclitaxel-coated balloons (PCBs) for in-stent restenosis (ISR) in SVs is limited.


OBJECTIVES - The aim of this study was to evaluate the efficacy and safety of PCB vs uncoated balloon angioplasty according to vessel size.


METHODS - AGENT IDE (A Clinical Trial to Assess the Agent Paclitaxel Coated PTCA Balloon Catheter for the Treatment of Subjects With In-Stent Restenosis) randomized 600 patients with ISR to treatment with PCBs or uncoated balloons (2:1). This prespecified analysis evaluated the treatment effect of PCBs in SV (reference vessel diameter [RVD] ≤2.75 mm) and large vessel (RVD >2.75 mm) ISR. The primary endpoint of 1-year target lesion failure (TLF) was a composite of target lesion revascularization, cardiac death, and target vessel-related myocardial infarction.


RESULTS -  Among 597 patients with known angiographic core laboratory-adjudicated vessel size, 56% had SVs (mean RVD 2.4 ± 0.3 mm) and 44% had large vessels (mean RVD 3.1 ± 0.3 mm). One-year TLF was 20.6% vs 22.6% in the SV vs large vessel groups, respectively (HR: 0.92; 95% CI: 0.65-1.31; P = 0.65). PCBs were associated with a 39% relative reduction in TLF compared with balloon angioplasty in patients with SVs (17.7% vs 27.4%; HR: 0.61; 95% CI: 0.37-0.99) and a 43% reduction in patients with large vessels (18.4% vs 30.5%; HR: 0.57; 95% CI: 0.34-0.96). The benefits of PCB use remained consistent, irrespective of vessel size (Pinteraction = 0.88). None of the patients treated with PCBs experienced definite or probable stent thrombosis.


CONCLUSIONS -  This prespecified subgroup analysis demonstrates that angioplasty with a PCB was associated with consistently reduced rates of 1-year TLF compared with an uncoated balloon in both SV and large vessel ISR patients. (A Clinical Trial to Assess the Agent Paclitaxel Coated PTCA Balloon Catheter for the Treatment of Subjects With In-Stent Restenosis [AGENT IDE]; NCT04647253).


Copyright © 2025 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.