Cardiovascular Magnetic Resonance as a complementary method to Transthoracic Echocardiography for Aortic Valve Area Estimation in patients with Aortic Stenosis: A systematic review and meta-analysis Comprehensive intravascular ultrasound assessment of stent area and its impact on restenosis and adverse cardiac events in 403 patients with unprotected left main disease Delirium After TAVR: Crosspassing the Limit of Resilience Reduced Leaflet Motion after Transcatheter Aortic-Valve Replacement Randomized Evaluation of Heart Failure With Preserved Ejection Fraction Patients With Acute Heart Failure and Dopamine - The ROPA-DOP Trial Active SB-P Versus Conventional Approach to the Protection of High-Risk Side Branches: The CIT-RESOLVE Trial Empagliflozin and Progression of Kidney Disease in Type 2 Diabetes Early Rhythm-Control Therapy in Patients with Atrial Fibrillation Why NOBLE and EXCEL Are Consistent With Each Other and With Previous Trials Impact of Positive and Negative Lesion Site Remodeling on Clinical Outcomes : Insights From PROSPECT

Original Research2017 Dec;30(6):564-569.

JOURNAL:J Interv Cardiol. Article Link

Diagnostic accuracy of instantaneous wave free-ratio in clinical practice

Ding WY, Nair S, Appleby C. Keywords: fractional flow reserve; functional testing; instantaneous wave-free ratio; pressure wire studies

ABSTRACT


AIMS - To evaluate the correlation between iFR and FFR in real-world clinical practice.


METHODS AND RESULTS - Retrospective, single-centre study of 229 consecutive pressure-wire studies (np  = 158). Real-time iFR and FFR measurements were performed for angiographically borderline stenoses. Functionally significant stenoses were defined as iFR <0.86 or FFR ≤0.80. An iFR between 0.86 and 0.93 was considered within the grey zone (Hybrid approach). Median iFR and FFR (IQR) were 0.92 (0.87-0.95) and 0.83 (0.76-0.89), respectively. Pearson's correlation coefficient was 0.75 (P < 0.001). Bland-Altman plot showed a mean difference between iFR and FFR that remained consistent throughout the range of values. The optimal iFR cutoff was 0.91-sensitivity 80%, specificity 82% with ROC area under curve of 89%. Using the Hybrid iFR-FFR strategy, we demonstrated high accuracy of iFR results-sensitivity 95%, specificity 96%, PPV 95%, and NPV 96%. In addition, this method would have avoided adenosine in 56% of patients. Mean follow-up period was 17.2 (±3.4) months. All-cause mortality was 3.2% (np = 5) and repeat intervention was required in six lesions (2.6%).


CONCLUSIONS - This study demonstrates that iFR is a valuable adjunct to FFR using the Hybrid iFR-FFR strategy in a real-world population. The use of adenosine may be avoided in about half the cases.