Relationship Between Hospital Surgical Aortic Valve Replacement Volume and Transcatheter Aortic Valve Replacement Outcomes Effects of Icosapent Ethyl on Total Ischemic Events: From REDUCE-IT Higher neutrophil-to-lymphocyte ratio (NLR) increases the risk of suboptimal platelet inhibition and major cardiovascular ischemic events among ACS patients receiving dual antiplatelet therapy with ticagrelor Ten-Year All-Cause Death According to Completeness of Revascularization in Patients With Three-Vessel Disease or Left Main Coronary Artery Disease: Insights From the SYNTAX Extended Survival Study Transcatheter Aortic Valve Replacement: Role of Multimodality Imaging in Common and Complex Clinical Scenarios Short Length of Stay After Elective Transfemoral Transcatheter Aortic Valve Replacement Is Not Associated With Increased Early or Late Readmission Risk Myocardial bridging of the left anterior descending coronary artery is associated with reduced myocardial perfusion reserve: a 13N-ammonia PET study Left Ventricular Hypertrophy and Clinical Outcomes Over 5 Years After TAVR: An Analysis of the PARTNER Trials and Registries Coronary Protection to Prevent Coronary Obstruction During TAVR: A Multicenter International Registry Association of Coronary Artery Calcium With Long-term, Cause-Specific Mortality Among Young Adults

Original Research2017 Dec;30(6):564-569.

JOURNAL:J Interv Cardiol. Article Link

Diagnostic accuracy of instantaneous wave free-ratio in clinical practice

Ding WY, Nair S, Appleby C. Keywords: fractional flow reserve; functional testing; instantaneous wave-free ratio; pressure wire studies

ABSTRACT


AIMS - To evaluate the correlation between iFR and FFR in real-world clinical practice.


METHODS AND RESULTS - Retrospective, single-centre study of 229 consecutive pressure-wire studies (np  = 158). Real-time iFR and FFR measurements were performed for angiographically borderline stenoses. Functionally significant stenoses were defined as iFR <0.86 or FFR ≤0.80. An iFR between 0.86 and 0.93 was considered within the grey zone (Hybrid approach). Median iFR and FFR (IQR) were 0.92 (0.87-0.95) and 0.83 (0.76-0.89), respectively. Pearson's correlation coefficient was 0.75 (P < 0.001). Bland-Altman plot showed a mean difference between iFR and FFR that remained consistent throughout the range of values. The optimal iFR cutoff was 0.91-sensitivity 80%, specificity 82% with ROC area under curve of 89%. Using the Hybrid iFR-FFR strategy, we demonstrated high accuracy of iFR results-sensitivity 95%, specificity 96%, PPV 95%, and NPV 96%. In addition, this method would have avoided adenosine in 56% of patients. Mean follow-up period was 17.2 (±3.4) months. All-cause mortality was 3.2% (np = 5) and repeat intervention was required in six lesions (2.6%).


CONCLUSIONS - This study demonstrates that iFR is a valuable adjunct to FFR using the Hybrid iFR-FFR strategy in a real-world population. The use of adenosine may be avoided in about half the cases.


© 2017, Wiley Periodicals, Inc.