中国肺动脉高压诊断与治疗指南(2021版) Transcatheter Aortic Valve Replacement: Role of Multimodality Imaging in Common and Complex Clinical Scenarios MINOCA: a heterogenous group of conditions associated with myocardial damage Short Length of Stay After Elective Transfemoral Transcatheter Aortic Valve Replacement Is Not Associated With Increased Early or Late Readmission Risk Evolution of antithrombotic therapy in patients undergoing percutaneous coronary intervention: a 40-year journey Change in Kidney Function and 2-Year Mortality After Transcatheter Aortic Valve Replacement von Willebrand Factor and Management of Heart Valve Disease: JACC Review Topic of the Week Glycemic Index, Glycemic Load, and Cardiovascular Disease and Mortality Association of Sustained Blood Pressure Control with Multimorbidity Progression Among Older Adults 6-month versus 12-month or longer dual antiplatelet therapy after percutaneous coronary intervention in patients with acute coronary syndrome (SMART-DATE): a randomised, open-label, non-inferiority trial

Clinical Trial2004 Jun 2;43(11):1959-63.

JOURNAL:J Am Coll Cardiol. Article Link

Impact of final stent dimensions on long-term results following sirolimus-eluting stent implantation: serial intravascular ultrasound analysis from the sirius trial

Sonoda S, Morino Y, Ako J et al. Keywords: BMS; IVUS; MLA; MSA; SES

ABSTRACT


OBJECTIVES We assessed the predictive value of minimum stent area (MSA) for long-term patency of sirolimus-eluting stents (SES) implantation compared to bare metal stents (BMS).


BACKGROUND - Although MSA is a consistent predictor of in-stent restenosis, its predictive value in BMS is still limited because of biologic variability in the restenosis process.

METHODS - From the SIRolImUS (SIRIUS) trial, 122 cases (SES: 72; BMS: 50) with complete serial intravascular ultrasound (IVUS) (baseline and 8-month follow-up) were analyzed. Postprocedure MSA and follow-up minimum lumen area (MLA) were obtained. Based on previous physiologic studies, adequate stent patency at follow-up was defined as MLA >4 mm(2).

RESULTS - In both groups, a significant positive correlation was observed between baseline MSA and follow-up MLA (SES: p < 0.0001, BMS: p < 0.0001). However, SES showed higher correlation than BMS (0.8 vs. 0.65) with a higher regression coefficient (0.92 vs. 0.59). The sensitivity and specificity curves identified different optimal thresholds of MSA to predict adequate follow-up MLA: 5 mm(2) for SES and 6.5 mm(2) for BMS. The positive predictive values with these cutoff points were 90% and 56%, respectively.

CONCLUSIONS - In this SIRIUS IVUS substudy, SES reduced both biologic variability and restenosis, resulting in increased predictability of long-term stent patency with postprocedure MSA. In addition, SES had a considerably lower optimal MSA threshold compared to BMS.