Long-Term Outcomes of Patients With Mediastinal Radiation–Associated Coronary Artery Disease Undergoing Coronary Revascularization With Percutaneous Coronary Intervention and Coronary Artery Bypass Grafting Systematic Review and Network Meta‐Analysis Comparing Bifurcation Techniques for Percutaneous Coronary Intervention Current Status and Future Prospects of Transcatheter Mitral Valve Replacement: JACC State-of-the-Art Review Does pulsed field ablation regress over time? A quantitative temporal analysis of pulmonary vein isolation Implications of the local hemodynamic forces on the formation and destabilization of neoatherosclerotic lesions Transseptal puncture versus patent foramen ovale or atrial septal defect access for left atrial appendage closure 2015 ACC/HRS/SCAI Left Atrial Appendage Occlusion Device Societal Overview Evolving insights into the role of local shear stress in late stent failure from neoatherosclerosis formation and plaque destabilization Management and outcomes of patients with left atrial appendage thrombus prior to percutaneous closure Alcohol consumption, cardiac biomarkers, and risk of atrial fibrillation and adverse outcomes

Clinical Trial2004 Jun 2;43(11):1959-63.

JOURNAL:J Am Coll Cardiol. Article Link

Impact of final stent dimensions on long-term results following sirolimus-eluting stent implantation: serial intravascular ultrasound analysis from the sirius trial

Sonoda S, Morino Y, Ako J et al. Keywords: BMS; IVUS; MLA; MSA; SES

ABSTRACT


OBJECTIVES We assessed the predictive value of minimum stent area (MSA) for long-term patency of sirolimus-eluting stents (SES) implantation compared to bare metal stents (BMS).


BACKGROUND - Although MSA is a consistent predictor of in-stent restenosis, its predictive value in BMS is still limited because of biologic variability in the restenosis process.

METHODS - From the SIRolImUS (SIRIUS) trial, 122 cases (SES: 72; BMS: 50) with complete serial intravascular ultrasound (IVUS) (baseline and 8-month follow-up) were analyzed. Postprocedure MSA and follow-up minimum lumen area (MLA) were obtained. Based on previous physiologic studies, adequate stent patency at follow-up was defined as MLA >4 mm(2).

RESULTS - In both groups, a significant positive correlation was observed between baseline MSA and follow-up MLA (SES: p < 0.0001, BMS: p < 0.0001). However, SES showed higher correlation than BMS (0.8 vs. 0.65) with a higher regression coefficient (0.92 vs. 0.59). The sensitivity and specificity curves identified different optimal thresholds of MSA to predict adequate follow-up MLA: 5 mm(2) for SES and 6.5 mm(2) for BMS. The positive predictive values with these cutoff points were 90% and 56%, respectively.

CONCLUSIONS - In this SIRIUS IVUS substudy, SES reduced both biologic variability and restenosis, resulting in increased predictability of long-term stent patency with postprocedure MSA. In addition, SES had a considerably lower optimal MSA threshold compared to BMS.