Bioprosthetic valve oversizing is associated with increased risk of valve thrombosis following TAVR Is Cardiac Diastolic Dysfunction a Part of Post-Menopausal Syndrome? Suture- or Plug-Based Large-Bore Arteriotomy Closure: A Pilot Randomized Controlled Trial From Subclinical Atherosclerosis to Plaque Progression and Acute Coronary Events Clinical Risk Factors and Atherosclerotic Plaque Extent to Define Risk for Major Events in Patients Without Obstructive Coronary Artery Disease: The Long-Term Coronary Computed Tomography Angiography CONFIRM Registry Major Bleeding Rates in Atrial Fibrillation Patients on Single, Dual, or Triple Antithrombotic Therapy 2019 ACC/AHA/ASE Advanced Training Statement on Echocardiography (Revision of the 2003 ACC/AHA Clinical Competence Statement on Echocardiography): A Report of the ACC Competency Management Committee Dual Antiplatelet Therapy Duration: Reconciling the Inconsistencies The Burden of Cardiovascular Diseases Among US States, 1990-2016 Haemodynamic-guided management of heart failure (GUIDE-HF): a randomised controlled trial

Clinical Trial2004 Jun 2;43(11):1959-63.

JOURNAL:J Am Coll Cardiol. Article Link

Impact of final stent dimensions on long-term results following sirolimus-eluting stent implantation: serial intravascular ultrasound analysis from the sirius trial

Sonoda S, Morino Y, Ako J et al. Keywords: BMS; IVUS; MLA; MSA; SES

ABSTRACT


OBJECTIVES We assessed the predictive value of minimum stent area (MSA) for long-term patency of sirolimus-eluting stents (SES) implantation compared to bare metal stents (BMS).


BACKGROUND - Although MSA is a consistent predictor of in-stent restenosis, its predictive value in BMS is still limited because of biologic variability in the restenosis process.

METHODS - From the SIRolImUS (SIRIUS) trial, 122 cases (SES: 72; BMS: 50) with complete serial intravascular ultrasound (IVUS) (baseline and 8-month follow-up) were analyzed. Postprocedure MSA and follow-up minimum lumen area (MLA) were obtained. Based on previous physiologic studies, adequate stent patency at follow-up was defined as MLA >4 mm(2).

RESULTS - In both groups, a significant positive correlation was observed between baseline MSA and follow-up MLA (SES: p < 0.0001, BMS: p < 0.0001). However, SES showed higher correlation than BMS (0.8 vs. 0.65) with a higher regression coefficient (0.92 vs. 0.59). The sensitivity and specificity curves identified different optimal thresholds of MSA to predict adequate follow-up MLA: 5 mm(2) for SES and 6.5 mm(2) for BMS. The positive predictive values with these cutoff points were 90% and 56%, respectively.

CONCLUSIONS - In this SIRIUS IVUS substudy, SES reduced both biologic variability and restenosis, resulting in increased predictability of long-term stent patency with postprocedure MSA. In addition, SES had a considerably lower optimal MSA threshold compared to BMS.