Outcomes with drug-coated balloons in small-vessel coronary artery disease Chronic Total Occlusion Interventions: Update on Current Tips and Tricks Bare metal or drug-eluting stent versus drug-coated balloon in non-ST-elevation myocardial infarction: the randomised PEPCAD NSTEMI trial Treating Bifurcation Lesions: The Result Overcomes the Technique Multicenter Registry of Real-World Patients With Severely Calcified Coronary Lesions Undergoing Orbital Atherectomy: 1-Year Outcomes Impact of stent deformity induced by the kissing balloon technique for bifurcating lesions on in-stent restenosis after coronary intervention The Hybrid Approach to Chronic Total Occlusion Percutaneous Coronary Intervention: Update From the PROGRESS CTO Registry Applications of left ventricular strain measurements to patients undergoing chemotherapy Percutaneous coronary intervention with drug-coated balloon-only strategy in stable coronary artery disease and in acute coronary syndromes: An all-comers registry study Percutaneous coronary interventional strategies for treatment of in-stent restenosis: a network meta-analysis

Clinical Trial2013 Mar;6(3):228-36.

JOURNAL:JACC Cardiovasc Interv. Article Link

Optical coherence tomography versus intravascular ultrasound to evaluate coronary artery disease and percutaneous coronary intervention

Bezerra HG, Attizzani GF, Sirbu V et al. Keywords: IVUS; OCT; PCI

ABSTRACT


OBJECTIVESWe compared intravascular ultrasound (IVUS) and 2 different generations of optical coherence tomography (OCT)-time-domain OCT (TD-OCT) and frequency-domain OCT (FD-OCT)-for the assessment of coronary disease and percutaneous coronary intervention (PCI) using stents.


BACKGROUND - OCT is a promising light-based intravascular imaging modality with higher resolution than IVUS. However, the paucity of data on OCT image quantification has limited its application in clinical practice.

METHODS - A total of 227 matched OCT and IVUS pull backs were studied. One hundred FD-OCT and IVUS pull backs in nonstented (n = 56) and stented (n = 44) vessels were compared. Additionally, 127 matched TD-OCT and IVUS images were compared in stented vessels.

RESULTS - FD-OCT depicted more severe native coronary disease than IVUS; minimal lumen area (MLA) was 2.33 ± 1.56 mm(2) versus 3.32 ± 1.92 mm(2), respectively (p < 0.001). Reference vessel dimensions were equivalent between FD-OCT and IVUS in both native and stented coronaries, but TD-OCT detected smaller reference lumen size compared with IVUS. Immediately post-PCI, in-stent MLAs were similar between FD-OCT and IVUS, but at follow-up, both FD-OCT and TD-OCT detected smaller MLAs than did IVUS, likely due to better detection of neointimal hyperplasia (NIH). Post-PCI malapposition and tissue prolapse were more frequently identified by FD-OCT.

CONCLUSIONS - FD-OCT generates similar reference lumen dimensions but higher degrees of disease severity and NIH, as well as better detection of malapposition and tissue prolapse compared with IVUS. First-generation TD-OCT was associated with smaller reference vessel dimensions compared with IVUS.

Copyright © 2013 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.