Treating Multivessel Coronary Artery Disease in ST-Segment Elevation Myocardial Infarction: Why, How, and When? Influence of LDL-Cholesterol Lowering on Cardiovascular Outcomes in Patients With Diabetes Mellitus Undergoing Coronary Revascularization Safety and feasibility of robotic percutaneous coronary intervention: PRECISE (Percutaneous Robotically-Enhanced Coronary Intervention) Study The optimal duration of dual antiplatelet therapy after coronary stent implantation: to go too far is as bad as to fall short Impact of Statins on Cardiovascular Outcomes Following Coronary Artery Calcium Scoring Utilization and programming of an automatic MRI recognition feature for cardiac rhythm management devices Causes, Timing, and Impact of Dual Antiplatelet Therapy Interruption for Surgery (from the Patterns of Non-adherence to Anti-platelet Regimens In Stented Patients Registry) Association of CYP2C19 Loss-of-Function Alleles with Major Adverse Cardiovascular Events of Clopidogrel in Stable Coronary Artery Disease Patients Undergoing Percutaneous Coronary Intervention: Meta-analysis Incidence, Predictors, and Outcomes of In-Hospital Percutaneous Coronary Intervention Following Coronary Artery Bypass Grafting Cardiorespiratory Fitness and Mortality in Healthy Men and Women

Clinical Trial2011 Apr;6(9):1037-45.

JOURNAL:EuroIntervention. Article Link

First-in-man evaluation of intravascular optical frequency domain imaging (OFDI) of Terumo: a comparison with intravascular ultrasound and quantitative coronary angiography

Okamura T, Onuma Y, Garcia-Garcia HM et al. Keywords: IVUS; stents; OCT; coronary imaging; optical frequency domain imaging

ABSTRACT


AIMS The objective of this study is to evaluate the feasibility and safety of imaging human coronary arteries in vivo by optical frequency domain imaging (OFDI) in comparison to intravascular ultrasound (IVUS). OFDI has been recently developed to overcome the limitations of conventional time-domain optical coherence tomography (OCT), namely the need for proximal balloon occlusion. The Terumo-OFDI system is capable of acquiring images with high-speed automated pullback (up to 40 mm/sec) and requires only a short injection (3-4 sec) of small amount of x-ray contrast (9-16 ml).


METHODS AND RESULTS - Nineteen patients who underwent stent implantation were enrolled. IVUS/OFDI were performed before and after stenting. The incidences of any adverse event and angiographic adverse findings were recorded. Lumen area (LA) was measured by IVUS and OFDI at 1 mm intervals in the stented segments (n=19) as well as in the proximal, distal, and to-be-stented segments (n=40). In addition, lumen area in the stented segment was also measured by edge (E-) and video-densitometric (VD-) quantitative coronary angiography (QCA). The OFDI images were obtained without any adverse event related to imaging procedures. Post stenting (n=19), minimal LA (MLA) measured by OFDI (5.84 ± 1.89 mm2) was larger than that of E-QCA (4.16 ± 1.46 mm2, p<0.001) and VD-QCA (4.92 ± 1.55 mm2, p<0.05). It was smaller than IVUS-MLA (6.26 ± 2.01 mm2, N.S.) but the correlation between the two measurements was highly significant (R2=0.82, p<0.001).

CONCLUSIONS - The OFDI imaging is feasible both before and after stenting and has a promising safety profile. The OFDI provided clear high resolution images and robust lumen measurements.