CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

Impact of percutaneous coronary intervention extent, complexity and platelet reactivity on outcomes after drug-eluting stent implantation Association Between Living in Food Deserts and Cardiovascular Risk Percutaneous Intervention for Concurrent Chronic Total Occlusions in Patients With STEMI: The EXPLORE Trial Effects of clopidogrel vs. prasugrel vs. ticagrelor on endothelial function, inflammatory parameters, and platelet function in patients with acute coronary syndrome undergoing coronary artery stenting: a randomized, blinded, parallel study Invasive Versus Medical Management in Patients With Prior Coronary Artery Bypass Surgery With a Non-ST Segment Elevation Acute Coronary Syndrome: A Pilot Randomized Controlled Trial Application of High-Sensitivity Troponin in Suspected Myocardial Infarction Ejection Fraction Pros and Cons: JACC State-of-the-Art Review 2016 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure No causal effects of plasma homocysteine levels on the risk of coronary heart disease or acute myocardial infarction: A Mendelian randomization study Interleukin-1 Beta as a Target for Atherosclerosis Therapy: Biological Basis of CANTOS and Beyond

Clinical TrialVolume 11, Issue 10, May 2018

JOURNAL:JACC Cardiovasc Interv. Article Link

Comparison of a Novel Biodegradable Polymer Sirolimus-Eluting Stent With a Durable Polymer Everolimus-Eluting Stent 5-Year Outcomes of the Randomized BIOFLOW-II Trial

T Lefèvre, M Haude, FJ Neumann et al. Keywords: biodegradable polymer; coronary artery disease; drug-eluting stent(s); sirolimus

ABSTRACT


OBJECTIVES - The authors aimed to compare long-term data of an ultrathin cobalt-chromium stent with passive silicon carbide coating and an active biodegradable polymer that releases sirolimus (O-SES) (Orsiro, BIOTRONIK, Bülach, Switzerland) with the durable polymer-based Xience Prime everolimus-eluting stent (X-EES) (Abbott Vascular, Santa Clara, California).


BACKGOURND - Biodegradable polymer stents have been developed aiming to overcome long-term detrimental effects of durable polymer stents, ultimately leaving a bare-metal stent in the vessel.

METHODS - This multicenter, assessor-blinded trial randomized 452 patients with 505 lesions to either O-SES or X-EES in a 2:1 fashion. Endpoints at 5 years were target lesion failure (TLF), its components, and stent thrombosis.

RESULTS - TLF occurred in 10.4% (n = 30) of O-SES patients versus 12.7% (n = 19) of X-EES patients (p = 0.473), overall stent thrombosis occurred in 0.7% (n = 2) versus 2.8% (n = 4) (p = 0.088), and definite stent thrombosis in 0% versus 0.7% (n = 1) (p = 0.341). Post hoc analysis was performed in diabetic patients (n = 128) and vessels ≤2.75 mm (n = 259). In diabetic patients, the O-SES group had numerically more target lesion revascularizations (13.5% vs. 4.5%; p = 0.138), but fewer cardiac deaths (1.3% vs. 6.9%; p = 0.089) and stent thrombosis (0% vs. 6.9%; p = 0.039). In small vessels, the O-SES group had a significantly lower 5-year mortality (3.7% vs. 11.3%; p = 0.022).

CONCLUSIONS - At 5 years, the biodegradable polymer O-SES demonstrated low TLF rates comparable to the durable polymer X-EES, confirming its long-term safety and performance. Particularly encouraging is the absence of definite stent thrombosis.