Bypass Surgery or Stenting for Left Main Coronary Artery Disease in Patients With Diabetes Effect of Intravascular Ultrasound-Guided Drug-Eluting Stent Implantation: Five-Year Follow-Up of the IVUS-XPL Randomized Trial Positive remodelling of coronary arteries on computed tomography coronary angiogram: an observational study Intravascular ultrasound guidance to minimize the use of iodine contrast in percutaneous coronary intervention: the MOZART (Minimizing cOntrast utiliZation With IVUS Guidance in coRonary angioplasTy) randomized controlled trial Left Main Revascularization With PCI or CABG in Patients With Chronic Kidney Disease: EXCEL Trial Therapeutic Options for In-Stent Restenosis Outcomes After Left Main Percutaneous Coronary Intervention Versus Coronary Artery Bypass Grafting According to Lesion Site Results From the EXCEL Trial The Utility of Contrast Medium Fractional Flow Reserve in Functional Assessment Of Coronary Disease in Daily Practice In vivo intravascular ultrasound-derived thin-cap fibroatheroma detection using ultrasound radiofrequency data analysis Diagnosis of Nonischemic Stage B Heart Failure in Type 2 Diabetes Mellitus: Optimal Parameters for Prediction of Heart Failure

Clinical TrialVolume 11, Issue 10, May 2018

JOURNAL:JACC Cardiovasc Interv. Article Link

Comparison of a Novel Biodegradable Polymer Sirolimus-Eluting Stent With a Durable Polymer Everolimus-Eluting Stent 5-Year Outcomes of the Randomized BIOFLOW-II Trial

T Lefèvre, M Haude, FJ Neumann et al. Keywords: biodegradable polymer; coronary artery disease; drug-eluting stent(s); sirolimus

ABSTRACT


OBJECTIVES - The authors aimed to compare long-term data of an ultrathin cobalt-chromium stent with passive silicon carbide coating and an active biodegradable polymer that releases sirolimus (O-SES) (Orsiro, BIOTRONIK, Bülach, Switzerland) with the durable polymer-based Xience Prime everolimus-eluting stent (X-EES) (Abbott Vascular, Santa Clara, California).


BACKGOURND - Biodegradable polymer stents have been developed aiming to overcome long-term detrimental effects of durable polymer stents, ultimately leaving a bare-metal stent in the vessel.

METHODS - This multicenter, assessor-blinded trial randomized 452 patients with 505 lesions to either O-SES or X-EES in a 2:1 fashion. Endpoints at 5 years were target lesion failure (TLF), its components, and stent thrombosis.

RESULTS - TLF occurred in 10.4% (n = 30) of O-SES patients versus 12.7% (n = 19) of X-EES patients (p = 0.473), overall stent thrombosis occurred in 0.7% (n = 2) versus 2.8% (n = 4) (p = 0.088), and definite stent thrombosis in 0% versus 0.7% (n = 1) (p = 0.341). Post hoc analysis was performed in diabetic patients (n = 128) and vessels ≤2.75 mm (n = 259). In diabetic patients, the O-SES group had numerically more target lesion revascularizations (13.5% vs. 4.5%; p = 0.138), but fewer cardiac deaths (1.3% vs. 6.9%; p = 0.089) and stent thrombosis (0% vs. 6.9%; p = 0.039). In small vessels, the O-SES group had a significantly lower 5-year mortality (3.7% vs. 11.3%; p = 0.022).

CONCLUSIONS - At 5 years, the biodegradable polymer O-SES demonstrated low TLF rates comparable to the durable polymer X-EES, confirming its long-term safety and performance. Particularly encouraging is the absence of definite stent thrombosis.