Ranolazine in High-Risk Patients With Implanted Cardioverter-Defibrillators - The RAID Trial Complete Versus Culprit-Only Revascularization in STEMI: a Contemporary Review Intraluminal Intensity of Blood Speckle on Intravascular Ultrasound, a Novel Predictor of Periprocedural Myocardial Injury After Coronary Stenting Coronary Catheterization and Percutaneous Coronary Intervention in China: 10-Year Results From the China PEACE-Retrospective CathPCI Study Prevention, Diagnosis, and Management of Radiation-Associated Cardiac Disease: JACC Scientific Expert Panel Acute Myocardial Infarction after Laboratory-Confirmed Influenza Infection Effect of Shorter Door-to-Balloon Times Over 20 Years on Outcomes of Patients With Anterior ST-Elevation Myocardial Infarction Undergoing Primary Percutaneous Coronary Intervention Prognostic significance of QRS fragmentation and correlation with infarct size in patients with anterior ST-segment elevation myocardial infarction treated with percutaneous coronary intervention: Insights from the INFUSE-AMI trial 2018 AHA/ACC/AACVPR/AAPA/ABC/ACPM/ADA/AGS/APhA/ASPC/NLA/PCNA Guideline on the Management of Blood Cholesterol: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines A case of influenza type a myocarditis that presents with ST elevation MI, cardiogenic shock, acute renal failure, and rhabdomyolysis and with rapid recovery after treatment with oseltamivir and intra-aortic balloon pump support

Clinical Trial2018 Mar;11(3):e005912.

JOURNAL:Circ Arrhythm Electrophysiol. Article Link

Can We Use the Intrinsic Left Ventricular Delay (QLV) to Optimize the Pacing Configuration for Cardiac Resynchronization Therapy With a Quadripolar Left Ventricular Lead?

van Everdingen WM, Zweerink A, Cramer MJ et al. Keywords: cardiac resynchronization therapy; heart failure; hemodynamics; humans; stroke volume

ABSTRACT


BACKGROUNDPrevious studies indicated the importance of the intrinsic left ventricular (LV) electric delay (QLV) for optimal benefit to cardiac resynchronization therapy. We investigated the use of QLV for achieving optimal acute hemodynamic response to cardiac resynchronization therapy with a quadripolar LV lead.


METHODS AND RESULTS - Forty-eight heart failure patients with a left bundle branch block were prospectively enrolled (31 men; age, 66±10 years; LV ejection fraction, 28±8%; QRS duration, 176±14 ms). Immediately after cardiac resynchronization therapy implantation, invasive LV pressure-volume loops were recorded during biventricular pacing with each separate electrode at 4 atrioventricular delays. Acute cardiac resynchronization therapy response, measured as change in stroke work (Δ%SW) compared with intrinsic conduction, was related to intrinsic interval between Q on the ECG and LV sensing delay (QLV), normalized for QRS duration (QLV/QRSd), and electrode position. QLV/QRSd was 84±9% and variation between the 4 electrodes 9±5%. Δ%SW was 89±64% and varied by 39±36% between the electrodes. In univariate analysis, an anterolateral or lateral electrode position and a high QLV/QRSd had a significant association with a large Δ%SW (all P <0.01). In a combined model, only QLV/QRSd remained significantly associated with Δ%SW (P<0.05). However, a direct relation between QLV/QRSd and Δ%SW was only seen in 24 patients, whereas 24 patients showed an inverse relation.

CONCLUSIONS - The large variation in acute hemodynamic response indicates that the choice of the stimulated electrode on a quadripolar lead is important. Although QLV/QRSd was associated with acute hemodynamic response at group level, it cannot be used to select the optimal electrode in the individual patient.