CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

High-Sensitivity Troponin and The Application of Risk Stratification Thresholds in Patients with Suspected Acute Coronary Syndrome Multivessel PCI Guided by FFR or Angiography for Myocardial Infarction Deficiency of GATA3-Positive Macrophages Improves Cardiac Function Following Myocardial Infarction or Pressure Overload Hypertrophy Open sesame technique in percutaneous coronary intervention for ST-elevation myocardial infarction Optimal medical therapy vs. coronary revascularization for patients presenting with chronic total occlusion: A meta-analysis of randomized controlled trials and propensity score adjusted studies Fractional flow reserve vs. angiography in guiding management to optimize outcomes in non-ST-segment elevation myocardial infarction: the British Heart Foundation FAMOUS-NSTEMI randomized trial The year in cardiovascular medicine 2020: acute coronary syndromes and intensive cardiac care Clinical and genetic characteristics of pulmonary arterial hypertension in Lebanon Imaging Coronary Anatomy and Reducing Myocardial Infarction 2017 AHA/ACC Clinical Performance and Quality Measures for Adults With ST-Elevation and Non–ST-Elevation Myocardial Infarction: A Report of the American College of Cardiology/American Heart Association Task Force on Performance Measures

Clinical Trial2018 Mar;11(3):e005912.

JOURNAL:Circ Arrhythm Electrophysiol. Article Link

Can We Use the Intrinsic Left Ventricular Delay (QLV) to Optimize the Pacing Configuration for Cardiac Resynchronization Therapy With a Quadripolar Left Ventricular Lead?

van Everdingen WM, Zweerink A, Cramer MJ et al. Keywords: cardiac resynchronization therapy; heart failure; hemodynamics; humans; stroke volume

ABSTRACT


BACKGROUNDPrevious studies indicated the importance of the intrinsic left ventricular (LV) electric delay (QLV) for optimal benefit to cardiac resynchronization therapy. We investigated the use of QLV for achieving optimal acute hemodynamic response to cardiac resynchronization therapy with a quadripolar LV lead.


METHODS AND RESULTS - Forty-eight heart failure patients with a left bundle branch block were prospectively enrolled (31 men; age, 66±10 years; LV ejection fraction, 28±8%; QRS duration, 176±14 ms). Immediately after cardiac resynchronization therapy implantation, invasive LV pressure-volume loops were recorded during biventricular pacing with each separate electrode at 4 atrioventricular delays. Acute cardiac resynchronization therapy response, measured as change in stroke work (Δ%SW) compared with intrinsic conduction, was related to intrinsic interval between Q on the ECG and LV sensing delay (QLV), normalized for QRS duration (QLV/QRSd), and electrode position. QLV/QRSd was 84±9% and variation between the 4 electrodes 9±5%. Δ%SW was 89±64% and varied by 39±36% between the electrodes. In univariate analysis, an anterolateral or lateral electrode position and a high QLV/QRSd had a significant association with a large Δ%SW (all P <0.01). In a combined model, only QLV/QRSd remained significantly associated with Δ%SW (P<0.05). However, a direct relation between QLV/QRSd and Δ%SW was only seen in 24 patients, whereas 24 patients showed an inverse relation.

CONCLUSIONS - The large variation in acute hemodynamic response indicates that the choice of the stimulated electrode on a quadripolar lead is important. Although QLV/QRSd was associated with acute hemodynamic response at group level, it cannot be used to select the optimal electrode in the individual patient.

© 2018 American Heart Association, Inc.