CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

Reduced Leaflet Motion after Transcatheter Aortic-Valve Replacement Randomized Evaluation of Heart Failure With Preserved Ejection Fraction Patients With Acute Heart Failure and Dopamine - The ROPA-DOP Trial Study of Two Dose Regimens of Ticagrelor Compared with Clopidogrel in Patients Undergoing Percutaneous Coronary Intervention for Stable Coronary Artery Disease (STEEL-PCI) Why and How to Measure Aortic Valve Calcification in Patients With Aortic Stenosis Combined use of OCT and IVUS in spontaneous coronary artery dissection Primary Prevention Trial Designs Using Coronary Imaging: A National Heart, Lung, and Blood Institute Workshop Transcatheter Aortic Valve Replacement vs Surgical Replacement in Patients With Pure Aortic Insufficiency The Utility of Rapid Atrial Pacing Immediately Post-TAVR to Predict the Need for Pacemaker Implantation Use of IVUS guided coronary stenting with drug eluting stent: a systematic review and meta-analysis of randomized controlled clinical trials and high quality observational studies Comparison of plaque characteristics in narrowings with ST-elevation myocardial infarction (STEMI), non-STEMI/unstable angina pectoris and stable coronary artery disease (from the ADAPT-DES IVUS Substudy)

Clinical Trial2018 Jun 21. [Epub ahead of print]

JOURNAL:Eur Heart J. Article Link

Japan-United States of America Harmonized Assessment by Randomized Multicentre Study of OrbusNEich's Combo StEnt (Japan-USA HARMONEE) study: primary results of the pivotal registration study of combined endothelial progenitor cell capture and drug-eluting stent in patients with ischaemic coronary disease and non-ST-elevation acute coronary syndrome

Saito S, Krucoff MW, Nakamura S et al. Keywords: DES; RCT; PCI; Endothelial progenitor cells; Japan

ABSTRACT


AIMSHarmonized Assessment by Randomized Multicentre Study of OrbusNEich's Combo StEnt (HARMONEE) (NCT02073565) was a randomized pivotal registration trial of the Combo stent, which combined sirolimus and an abluminal bioabsorbable polymer with a novel endoluminal anti-CD34+ antibody coating designed to capture endothelial progenitor cells (EPC) and promote percutaneous coronary intervention (PCI) site healing.


METHODS AND RESULTS - Clinically stabilized PCI subjects were randomized 1:1 to receive Combo or everolimus-eluting stents (EES). Between February 2014 and June 2016, 572 subjects with 675 coronary lesions underwent 1-year angiography and fractional flow reserve, with optical coherence tomography (OCT) in the first 140 patients. The primary clinical endpoint was non-inferior 1-year target vessel failure (TVF). The primary mechanistic endpoint of EPC capture activity was superior strut coverage by OCT. Target vessel failure occurred in 7.0% Combo (20/287) vs. 4.2% EES (12/285), a 2.8% [95% confidence interval (95% CI) -1.0%, 6.5%] difference, meeting the non-inferiority hypothesis (P = 0.02). There were no cardiac deaths, with one stent thrombosis observed in the EES group. Quantitative coronary angiography late loss with Combo was equivalent to EES. Optical coherence tomography strut coverage at 1 year was superior with Combo vs. EES [91.3% (95% CI 88.7%, 93.8%) vs. 74.8% (95% CI 70.0%, 79.6%), P < 0.001], with homogeneous tissue in 81.2% vs. 68.8%, respectively.

CONCLUSION - Combo stent demonstrated non-inferior 1-year TVF and late loss in a randomized comparison to EES, with superior strut-based tissue coverage by OCT as a surrogate of EPC capture technology activity.