CBS 2019
CBSMD教育中心
中 文

IVUS Guidance

Abstract

Recommended Article

Impact of intravascular ultrasound guidance on long-term mortality in stenting for unprotected left main coronary artery stenosis Meta-analysis of outcomes after intravascular ultrasound-guided versus angiography-guided drug-eluting stent implantation in 26,503 patients enrolled in three randomized trials and 14 observational studies Effects of Intravascular Ultrasound-Guided Versus Angiography-Guided New-Generation Drug-Eluting Stent Implantation: Meta-Analysis With Individual Patient-Level Data From 2,345 Randomized Patients Comparison of paclitaxel-eluting stents (Taxus) and everolimus-eluting stents (Xience) in left main coronary artery disease with 3 years follow-up (from the ESTROFA-LM registry) Novel predictor of target vessel revascularization after coronary stent implantation: Intraluminal intensity of blood speckle on intravascular ultrasound Intravascular ultrasound-derived minimal lumen area criteria for functionally significant left main coronary artery stenosis A three-vessel virtual histology intravascular ultrasound analysis of frequency and distribution of thin-cap fibroatheromas in patients with acute coronary syndrome or stable angina pectoris Defining a new standard for IVUS optimized drug eluting stent implantation: the PRAVIO study

Original Research2014 Aug;174(8):1360-6.

JOURNAL:JAMA Intern Med. Article Link

Long-term survival in patients undergoing percutaneous interventions with or without intracoronary pressure wire guidance or intracoronary ultrasonographic imaging: a large cohort study

Fröhlich GM, Redwood S, Rakhit R et al. Keywords: IVUS guided PCI; FFR; DES; outcome; cohort

ABSTRACT



IMPORTANCE - Intracoronary pressure wire-derived measurements of fractional flow reserve (FFR) and intravascular ultrasonography (IVUS) provide functional and anatomical information that can be used to guide coronary stent implantation. Although these devices are widely used and recommended by guidelines, limited data exist about their effect on clinical end points.


OBJECTIVE - To determine the effect on long-term survival of using FFR and IVUS during percutaneous coronary intervention (PCI).


DESIGN AND SETTING - Cohort study based on the pan-London (United Kingdom) PCI registry. In total, 64,232 patients are included in this registry covering the London, England, area.


PARTICIPANTS - All patients (n = 41,688) who underwent elective or urgent PCI in National Health Service hospitals in London between January 1, 2004, and July 31, 2011, were included. Patients with ST-segment elevation myocardial infarction (n = 11,370) were excluded.


INTERVENTIONS - Patients underwent PCI guided by angiography (visual lesion assessment) alone, PCI guided by FFR, or IVUS-guided PCI.


MAIN OUTCOMES AND MEASURES - The primary end point was all-cause mortality at a median of 3.3 years.


RESULTS - Fractional flow reserve was used in 2767 patients (6.6%) and IVUS was used in 1831 patients (4.4%). No difference in mortality was observed between patients who underwent angiography-guided PCI compared with patients who underwent FFR-guided PCI (hazard ratio, 0.88; 95% CI, 0.67-1.16; P = .37). Patients who underwent IVUS had a slightly higher adjusted mortality (hazard ratio, 1.39; 95% CI, 1.09-1.78; P = .009) compared with patients who underwent angiography-guided PCI. However, this difference was no longer statistically significant in a propensity score-based analysis (hazard ratio, 1.33; 95% CI, 0.85-2.09; P = .25). The mean (SD) number of implanted stents was lower in the FFR group (1.1 [1.2] stents) compared with the IVUS group (1.6 [1.3]) and the angiography-guided group (1.7 [1.1]) (P < .001).


CONCLUSIONS AND RELEVANCE - In this large observational study, FFR-guided PCI and IVUS-guided PCI were not associated with improved long-term survival compared with standard angiography-guided PCI. The use of FFR was associated with the implantation of fewer stents.