CBS 2019
CBSMD教育中心
中 文

IVUS Guidance

Abstract

Recommended Article

First-in-man evaluation of intravascular optical frequency domain imaging (OFDI) of Terumo: a comparison with intravascular ultrasound and quantitative coronary angiography Assessment Of Proximal Left Anterior Descending Artery Size By Intravascular Ultrasound For Optimal Stent Sizing A volumetric intravascular ultrasound comparison of early drug-eluting stent thrombosis versus restenosis Utility of intravascular ultrasound guidance in patients undergoing percutaneous coronary intervention for type C lesions Angiographic and clinical comparisons of intravascular ultrasound- versus angiography-guided drug-eluting stent implantation for patients with chronic total occlusion lesions: two-year results from a randomised AIR-CTO study Intravascular ultrasound-guided percutaneous coronary intervention improves the clinical outcome in patients undergoing multiple overlapping drug-eluting stents implantation Temporal Trends in Inpatient Use of Intravascular Imaging Among Patients Undergoing Percutaneous Coronary Intervention in the United States The role of integrated backscatter intravascular ultrasound in characterizing bare metal and drug-eluting stent restenotic neointima as compared to optical coherence tomography

Original ResearchMarch, 2018 Volume 71, Issue 11 Supplement

JOURNAL:J Am Coll Cardiol. Article Link

Assessment Of Proximal Left Anterior Descending Artery Size By Intravascular Ultrasound For Optimal Stent Sizing

Shlofmitz E; Matsumura M; Mintz GS et al. Keywords: proximal left anterior descending artery; IVUS; stent sizing

ABSTRACT


BACKGROUND - As the LAD supplies almost half of the myocardium, the proximal LAD (PLAD) should rarely be small. Given the prognostic significance of both the PLAD and minimal stent area, we evaluated PLAD sizes by IVUS.

METHODS - From isolated review of the angiograms from 147 pts who underwent IVUS-guided stenting of de novo PLAD lesions, 4 interventional cardiologists (two of whom were experienced IVUS users) recommended the stent diameter. An IVUS core lab then analyzed the lesion and vessel segments. Based on the smallest mean IVUS vessel diameter (VD), the optimal stent diameter was chosen by downsizing by 0.25-0.5 mm, except in VD >4.0mm.

RESULTS - Mean age was 66 yrs, 30% had diabetes, and 44% presented with ACS. The proximal and distal VDs were 4.5 ± 0.6 mm and 4.0 ± 0.6 mm, respectively. The smallest IVUS VD was 3.9 ± 0.5 mm (occurring in the lesion and distal reference in 44% and 56% of cases). 2% of the smallest VDs were <3.0 mm, and 93% of IVUS-guided optimal stent diameters were ≥3.0 mm (Figure). The mean stent size recommended by the 4 ICs based on angiography was 3.2 ± 0.3 mm; stent size was underestimated by 58% and 65% of experienced and inexperienced IVUS users, respectively. In a logistic model, diabetes was the only predictor for underestimation (OR [95%CI]; 2.48 [1.25- 4.93], P=0.009).

CONCLUSION - Stent diameters in the PLAD are frequently under-estimated based on angiography alone, and should rarely by <3.0 mm. Irrespective of experience, routine IVUS use may result in more appropriate stent sizing in the PLAD.