A Meta-Analysis of Contemporary Lesion Modification Strategies During Percutaneous Coronary Intervention in 244,795 Patients From 22 Studies Drug-Coated Balloon Treatment for Femoropopliteal Artery Disease: The IN.PACT Global Study De Novo In-Stent Restenosis Imaging Cohort Coronary Artery Calcium Is Associated with Left Ventricular Diastolic Function Independent of Myocardial Ischemia Can the Vanishing Stent Reappear? Fix the Technique, or Fix the Device? Aggressive lipid-lowering therapy after percutaneous coronary intervention – for whom and how? Rotational Atherectomy Followed by Drug-Coated Balloon Dilation for Left Main In-Stent Restenosis in the Setting of Acute Coronary Syndrome Complicated with Right Coronary Chronic Total Occlusion SGLT2 Inhibitors in Patients With Heart Failure With Reduced Ejection Fraction: A Meta-Analysis of the EMPEROR-Reduced and DAPA-HF Trials Disrupting Fellow Education Through Group Texting: WhatsApp in Fellow Education? AIM2-driven inflammasome activation in heart failure Coronary Angiography after Cardiac Arrest — The Right Timing or the Right Patients?

Original Research2018 Aug;31(4):421-429.

JOURNAL:J Interv Cardiol. Article Link

Impact of stent deformity induced by the kissing balloon technique for bifurcating lesions on in-stent restenosis after coronary intervention

Gamou T, Sataka K,Yamagishi M et al. Keywords: coronary bifurcation; intravascular ultrasound; kissing balloon technique

ABSRACT



OBJECTIVES - To investigate the impact of stent deformity induced by final kissing balloon technique (KBT) for coronary bifurcation lesions on in-stent restenosis (ISR).

 

BACKGROUND - In experimental models, the detrimental effects of KBT have been clearly demonstrated, but few data exists regarding the impact of proximal stent deformity induced by KBT on clinical outcomes.

 

METHODS - We examined 370 coronary lesions where intravascular ultrasound (IVUS)-guided second-generation drug-eluting stent (DES) implantation for coronary bifurcation lesions was performed. Based on IVUS analysis, the stent symmetry index (minimum/maximum stent diameter) and stent overstretch index (the mean of stent diameter/the mean of reference diameter) were calculated in the proximal main vessel.

 

RESULTS - The stent symmetry index was significantly lower (0.75 ± 0.07 vs 0.88 ± 0.06, P < 0.0001) and the stent overstretch index was significantly higher (1.04 ± 0.08 vs 1.01 ± 0.06, P = 0.0007) in lesions with KBT (n = 174) compared to those without KBT (n = 196). The number of two-stent technique in lesions with KBT was 31 (18%). In multivariate analysis, the degree of stent deformity indices was not associated with ISR in lesions with KBT; however, two-stent technique use was the only independent predictor of ISR at 8 months (hazard ratio: 3.96, 95% confidence interval: 1.25-12.5, P = 0.01).

 

CONCLUSIONS - Second-generation DES deformity induced by KBT was not associated with mid-term ISR.

 

 

© 2018 Wiley Periodicals, Inc.