Original Research
2018 Aug;31(4):421-429.
JOURNAL:J Interv Cardiol.
Article Link

Impact of stent deformity induced by the kissing balloon technique for bifurcating lesions on in-stent restenosis after coronary intervention
Gamou T, Sataka K,Yamagishi M et al.

KEYWORDS
coronary bifurcation; intravascular ultrasound; kissing balloon technique



OBJECTIVES - To investigate the impact of stent deformity induced by final kissing balloon technique (KBT) for coronary bifurcation lesions on in-stent restenosis (ISR).

 

BACKGROUND - In experimental models, the detrimental effects of KBT have been clearly demonstrated, but few data exists regarding the impact of proximal stent deformity induced by KBT on clinical outcomes.

 

METHODS - We examined 370 coronary lesions where intravascular ultrasound (IVUS)-guided second-generation drug-eluting stent (DES) implantation for coronary bifurcation lesions was performed. Based on IVUS analysis, the stent symmetry index (minimum/maximum stent diameter) and stent overstretch index (the mean of stent diameter/the mean of reference diameter) were calculated in the proximal main vessel.

 

RESULTS - The stent symmetry index was significantly lower (0.75 ± 0.07 vs 0.88 ± 0.06, P < 0.0001) and the stent overstretch index was significantly higher (1.04 ± 0.08 vs 1.01 ± 0.06, P = 0.0007) in lesions with KBT (n = 174) compared to those without KBT (n = 196). The number of two-stent technique in lesions with KBT was 31 (18%). In multivariate analysis, the degree of stent deformity indices was not associated with ISR in lesions with KBT; however, two-stent technique use was the only independent predictor of ISR at 8 months (hazard ratio: 3.96, 95% confidence interval: 1.25-12.5, P = 0.01).

 

CONCLUSIONS - Second-generation DES deformity induced by KBT was not associated with mid-term ISR.

 

 

© 2018 Wiley Periodicals, Inc.