CBS 2019
CBSMD教育中心
English

左主干支架

科研文章

荐读文献

Left-main restenosis in the DES era-a call for action Current Interventions for the Left Main Bifurcation One or two stents for the distal Left Main bifurcation The DK crush V study - The DK crush V study Double Kissing Crush Versus Provisional Stenting for Left Main Distal Bifurcation Lesions: DKCRUSH-V Randomized Trial Comparison of double kissing crush versus Culotte stenting for unprotected distal left main bifurcation lesions: results from a multicenter, randomized, prospective DKCRUSH-III study Design and rationale for the treatment effects of provisional side branch stenting and DK crush stenting techniques in patients with unprotected distal left main coronary artery bifurcation lesions (DKCRUSH V) Trial Left Main Revascularization in 2017 Coronary Artery Bypass Grafting or Percutaneous Coronary Intervention? Stent fracture is associated with a higher mortality in patients with type-2 diabetes treated by implantation of a second-generation drug-eluting stent Usefulness of the SYNTAX score II to validate 2-year outcomes in patients with complex coronary artery disease undergoing percutaneous coronary intervention: A large single-center study Contemporary Approach to Coronary Bifurcation Lesion Treatment

Review Article18 July 2017, [Epub ahead of print]

JOURNAL:Lancet Article Link

2-year outcomes with the Absorb bioresorbable scaffold for treatment of coronary artery disease: a systematic review and meta-analysis of seven randomised trials with an individual patient data substudy

Z.A. Ali; P.W. Serruys; G.W. Stone et al Keywords: Absorb bioresorbable scaffold, individual patient

ABSTRACT


Background - Bioresorbable vascular scaffolds (BVS) offer the potential to improve long-term outcomes of percutaneous coronary intervention after their complete bioresorption. Randomised trials have shown non-inferiority between BVS and metallic drug-eluting stents at 1 year in composite safety and effectiveness outcomes, although some increases in rates of target vessel-related myocardial infarction and device thrombosis were identified. Outcomes of BVS following the first year after implantation are unknown. We sought to ascertain whether BVS are as safe and effective as drug-eluting stents within 2 years after implantation and between 1 and 2 years.


Methods - We did a systematic review and meta-analysis of randomised trials in which patients were randomly assigned to everolimus-eluting Absorb BVS or metallic everolimus-eluting stents (EES) and followed up for at least 2 years. We searched MEDLINE, the Cochrane database, TCTMD, ClinicalTrials.gov, Clinical Trial Results, CardioSource, and abstracts and presentations from major cardiovascular meetings up to April 1, 2017, to identify relevant studies. The primary efficacy outcome measure was the device-oriented composite endpoint (cardiac mortality, target vessel-related myocardial infarction, or ischaemia-driven target lesion revascularisation) and the primary safety outcome measure was definite or probable device thrombosis. Individual patient data from the four ABSORB trials were used for landmark and subgroup analysis and multivariable modelling.


Findings - We identified seven randomised trials in which 5583 patients were randomly assigned to Absorb BVS (n=3261) or metallic EES (n=2322) and followed up for 2 years. BVS had higher 2-year relative risks of the device-oriented composite endpoint than did EES (9.4% [304 of 3217] vs 7.4% [169 of 2299]; relative risk [RR] 1.29 [95% CI 1.08-1.56], p=0.0059). These differences were driven by increased rates of target vessel-related myocardial infarction (5.8% [187 of 3218] vs 3.2% [74 of 2299]; RR 1.68 [95% CI 1.29-2.19], p=0.0003) and ischaemia-driven target lesion revascularisation (5.3% [169 of 3217] vs 3.9% [90 of 2300]; 1.40 [1.09-1.80], p=0.0090) with BVS, with non-significant differences in cardiac mortality. The cumulative 2-year incidence of device thrombosis was higher with BVS than with EES (2.3% [73 of 3187] vs 0.7% [16 of 2281]; RR 3.35 [95% CI 1.96-5.72], p<0.0001). Landmark analysis between 1 and 2 years also showed higher rates of the device-oriented composite endpoint (3.3% [69 of 2100] vs 1.9% [23 of 1193]; RR 1.64 [95% CI 1.03-2.61], p=0.0376) and device thrombosis (0.5% [11 of 2085] vs none [0 of 1183], p<0.0001) in BVS-treated patients than in EES-treated patients.


Interpretation - BVS was associated with increased rates of composite device-oriented adverse events and device thrombosis cumulatively at 2 years and between 1 and 2 years of follow-up compared with EES.